Reliability and validity of the Biodex System 3 Pro Isokinetic Dynamometer velocity, torque and position measurements

UNCG Author/Contributor (non-UNCG co-authors, if there are any, appear on document)
David H. Perrin, Former Provost and Executive Vice Chancellor (Creator)
Sandra J. Shultz, Professor and Chair (Creator)
Institution
The University of North Carolina at Greensboro (UNCG )
Web Site: http://library.uncg.edu/

Abstract: This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5° increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30°/s to 500°/s) across a 70° arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300°/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.

Additional Information

Publication
European Journal of Applied Physiology. 91:22-29
Language: English
Date: 2004
Keywords
Muscle function, Muscle testing, Reliability, Validity