Reliability and validity of the Biodex System 3 Pro Isokinetic Dynamometer velocity, torque and position measurements
- UNCG Author/Contributor (non-UNCG co-authors, if there are any, appear on document)
- David H. Perrin, Former Provost and Executive Vice Chancellor (Creator)
- Sandra J. Shultz, Professor and Chair (Creator)
- Institution
- The University of North Carolina at Greensboro (UNCG )
- Web Site: http://library.uncg.edu/
Abstract: This study quantitatively assessed the mechanical reliability and validity of position, torque and velocity measurements of the Biodex System 3 isokinetic dynamometer. Trial-to-trial and day-to-day reliability were assessed during three trials on two separate days. To assess instrument validity, measurement of each variable using the Biodex System 3 dynamometer was compared to a criterion measure of position, torque and velocity. Position was assessed at 5° increments across the available range of motion of the dynamometer. Torque measures were assessed isometrically by hanging six different calibrated weights from the lever arm. Velocity was assessed (30°/s to 500°/s) across a 70° arc of motion by manually accelerating the weighted lever arm. With the exception of a systematic decrease in velocity at speeds of 300°/s and higher, the Biodex System 3 performed with acceptable mechanical reliability and validity on all variables tested.
Reliability and validity of the Biodex System 3 Pro Isokinetic Dynamometer velocity, torque and position measurements
PDF (Portable Document Format)
271 KB
Created on 1/1/2004
Views: 19547
Additional Information
- Publication
- European Journal of Applied Physiology. 91:22-29
- Language: English
- Date: 2004
- Keywords
- Muscle function, Muscle testing, Reliability, Validity