Calibration of Rotational Acceleration for the Rotarod Test of Rodent Motor Coordination
- UNCG Author/Contributor (non-UNCG co-authors, if there are any, appear on document)
- Douglas Wahlsten, Visiting Professor (Creator)
- Institution
- The University of North Carolina at Greensboro (UNCG )
- Web Site: http://library.uncg.edu/
Abstract: The latency of mice and rats to fall from the accelerating rotarod can differ markedly between laboratories using the same brand of rod as well as between studies using different kinds of rods. These discrepancies can arise from different rod diameters, surface textures, test protocols, or laboratory environmental factors beyond the test itself, but it is also possible that the actual acceleration rates of the different rods do not correspond to the nominal rates set on the devices. This paper describes a simple method to measure acceleration rate of the rotarod and to set the rate to a desired value for any brand of rod.
Calibration of Rotational Acceleration for the Rotarod Test of Rodent Motor Coordination
PDF (Portable Document Format)
641 KB
Created on 3/16/2011
Views: 9580
Additional Information
- Publication
- Journal of Neuroscience Methods, 178: 10-14.
- Language: English
- Date: 2009
- Keywords
- Behavioral test, Motor coordination, Task parameters, Test standardization, Mouse, Rat, Rotarod